Efficacy of smear layer removal from root canals using three agitation devices

root canalSandra Tipanta, Osvaldo Zmener and Cornelis Pameijer investigate the effectiveness of three different irrigation/agitation techniques (Endoactivator, Eddy and XP-endo Finisher) at removing debris and the smear layer from root canal systems.

Complete removal of debris (DE) and the smear layer (SL) from the root canal system is an essential step for endodontic success (Schafer and Zapke, 2000; Paque et al, 2011).

During root canal preparation, the operator relies foremost on endodontic instruments. However, an effective irrigation protocol is of critical importance. It cleans the root canal system from the SL. Then, flushes out DE and acts as a tissue solvent, bactericidal agent and lubricant.

The use of a sodium hypochlorite solution (NaOCl) removes loosely attached DE and organic material. Chelating agents such as ethylenediaminetetraacetic acid (EDTA) are necessary to effectively remove the inorganic part of the SL. (Hülsmann et al, 1997; Svec and Harrison 1977). Historically, different protocols have been recommended for root canal irrigation.

All have demonstrated that some amount of DE and the SL remained on the canal walls after instrumentation and irrigation. (Svec and Harrison, 1977; Haapasalo et al, 2014; Zehnder, 2006).

The use of a conventional needle irrigation technique is the traditional method of delivering irrigating solutions to the root canals. However, when using this technique, solutions do not always come into contact with all irregularities of the root canal walls (Schafer and Zapke, 2000; Paque et al, 2011; Blank-Gonçalves et al, 2011).

Techniques

To improve the efficacy of irrigation, agitation of the irrigating solution has been proposed (Blank-Gonçalves et al, 2011; Caron et al, 2010).

Endoactivator

Among these techniques, the Endoactivator (Dentsply Sirona) was found to improve, to some extent, the effectiveness of irrigation.

The Endoactivator is a cordless battery-activated sonic handpiece. It uses non-cutting polymer tips for a safe intracanal agitation of the irrigation solutions. The tips are available in three different sizes, producing 2,000 to 10,000 cycles per minute.

The instrument produces a three-dimensional movement with high amplitude acoustic oscillations (Gu et al, 2009). It was also recommended to use the Endoactivator after the cleaning and shaping of canals (Uroz-Torres et al, 2010).

Eddy

Eddy (VDW) is another sonic device that activates irrigating solutions by means of a 25/.04 smooth flexible non-cutting polyamide tip, powered by a sonic scaler at a frequency of 5,000 to 6,000 cycles per minute.

Eddy (EPt) produces an acoustic streaming which effectively removes DE and the SL that adheres to root canal walls, including in difficult-to-access areas of the canal system (Khaord et al, 2015; Urban et al, 2017; Bayari et al, 2017).

XP-endo Finisher

More recently, the XP-endo Finisher (FKG Dentaire SA) has also been introduced for the activation of irrigants.

The XP-endo Finisher (XPe) consists of a size 25/.00 non-tapered instrument. It is constructed of Maxwire Martensite-austenite Electropolish-flex alloy (Bao et al, 2017). This is a special nickel titanium alloy.

It’s capable of producing an expanding movement when the instrument is used at body temperature (35°C – 37°C). Under these conditions, the instrument can adapt three-dimensionally to the canal anatomy.

To be effective, the XP-endo Finisher must be used at 800 rpm and, after canal preparation, to a size 25 or larger (Bao et al, 2017).

The purpose of this in vitro SEM study was to compare the efficacy of Endoactivator, Eddy and XP-endo Finisher. Compare them in the removal of DE and the SL from root canal walls as a final step in irrigation. Conventional needle irrigation technique served as the control.

The null hypothesis was that there would be no significant differences among the tested irrigation systems.  

Materials and methods

The protocol of this study was approved by the Ethics Committee of the Argentine Dental Association (protocol 0119/18). Forty (n=40) extracted human lower premolars with single straight root canals were used. They were stored at 4°C in 0.1% thymol solution before being treated. The teeth were decoronated to a standardised length of 18 mm.

After the removal of gross pulpal tissues, the working length (WL) was established by advancing a size 10 K file into the canal until just visible at the apex under a stereomicroscope (Carl Zeiss, Oberkochen, Germany) followed by subtracting 1mm. All teeth were assigned to four groups of 10 teeth each (n=10), using a stratified sampling method to ensure that all groups had teeth of similar average dimensions. This was accomplished by the use of the Minitab Statistical Analysis Package (Minitab 10.1, Minitab Inc. State College, PA, USA).

In all groups the canals were prepared with Protaper Next niti instruments (Dentsply, Tulsa Dental, Tulsa, OK), strictly according to the manufacturer’s recommendations. Biomechanical preparation of the apical part of the canals was considered complete when an instrument #X3 (equivalent to a size #30/.07 taper) could easily be inserted to the WL while maintaining apical patency. Throughout preparation, the canals were irrigated with 5 ml of 5.25% NaOCl before and after each instrument using a 30-guage Navitip needle (Ultradent Products Inc, South Jordan, UT) placed 1 mm short of the WL without binding. After preparation, the canals of each group were subjected to different final irrigation/agitation techniques. Group 1: EAc; 2: PTe;  3: XPe; 4: CNi.

To ensure retention of the irrigants in the canal space the root apex was sealed with cyanoacrylate (La Gotita, Buenos Aires, Argentina) taking care to prevent getting cyanoacrylate in the canal. The final irrigation/activation procedures were performed according to the following protocols:

Group 1. EAc (n=10)

2 ml of 5.25% NaOCl was injected with a #30-gauge Navitip needle (Ultradent Products Inc.) and the solution activated for 30 seconds with a #25/.04 non-cutting polymer tip powered to 10,000 cycles per minute. The tip was used with in-and-out movements of approximately 6-8 mm to the full WL.

After flushing with saline, the irrigation/activation procedure continued by filling the canals with 17% EDTA and agitation with EAc for another 30 seconds. The cycle of irrigation/activation with 5.25% NaOCl was repeated and the canals were finally flushed with 5 ml of 5.25% NaOCl following by rinsing with copious amounts of saline and dried with capillary suction tips and paper points. The polymer tip was discarded after each use.

Group 2. EPt (n=10)

The procedure was similar to that used in Group 1. However, the agitation was performed with EPt. Using a #20/.02 non-cutting polymer tip powered at a frequency of 6000 Hz by an air scaler (NSK AS2000; NSK, Japan) set at maximum speed. The Eddy tip was discarded after each use.

Group 3. XPe (n=10)

A similar procedure was used as in to Group 1 and 2, but agitation was performed with the XPe activated at 800 rpm in a torque controlled contra-angle. To control the proper length a rubber stop was used at a predetermined length. The XPe was also discarded after each use.

Group 4. CNi (n=10)

After instrumentation a #30-gauge Navitip needle was placed 1 mm short of the WL and the canal rinsed with 3 ml of 5.25% NaOCl. The irrigating solution was left in place for 30 seconds, evacuated and followed by 3 ml of 17% EDTA for 30 seconds. This cycle of irrigation was repeated and the canals were finally flushed with 5 ml of 5.25% NaOCl followed by rinsing with copious amounts of saline. They were then dried with capillary suction tips and paper points. The needle was used for one canal only.

Irrigation

Irrigation for all groups was performed in a temperature-controlled environment of 37°C. The access openings were sealed with Cavit (3M ESPE, Seefeld, Germany) and the samples coded to allow for a blinded evaluation. In preparation for scanning electron microscopy, a groove was prepared on the buccal and lingual root surface with a diamond disc under copious water cooling making sure not to penetrate the root canal space.

The samples were then immersed in liquid nitrogen and split longitudinally with a mallet and chisel. Note the seal of the access openings prevented the liquid nitrogen from penetrating the root canal.

Teeth with evidence that the groove had penetrated into the canal space or that exhibited an irregular cleavage were discarded and replaced with new specimens. The paired halves of each tooth were mounted side by side on an aluminium stub, coated with gold-palladium and examined in a scanning electron microscope (JEOL JSM 6490LV, Tokyo, Japan) operated at 15.5 Kv.

Serial SEM photomicrographs were made at 1000x magnification and aligned in such a manner that they generated a horizontal panoramic view covering the total circumference of the canal walls at levels of 1, 5 and 10 mm from the WL. Digital images were transferred to a computer and analysed using Image-Proplus 6.0 software (Media Cybernetics, Bethesda, MD).

Evaluation

Table 1. Mean (SD) scores of debris removal at 1, 5 and 10 mm from the WL.

For evaluation a slight modification of the method described by Mayer et al (2002) was used. Briefly, a 300-micron square grid was superimposed on the photographs and a determination was made as to the presence of DE and SL remnants. Each square was considered as an assessment unit and the examination space was set at 600 microns square.

The total area of each of the experimental levels was analysed by two evaluators who were blinded to the experimental groups and who independently scored the presence of DE and SL. Before the analysis both evaluators were calibrated by having them analyse a set of SEM photomicrographs at a magnification of 1000x obtained from the canal walls of teeth subjected to endodontic instrumentation and different irrigation regimens. Inter-observer reproducibility was measured by the Kappa coefficient.

The amount of DE and SL layer in each measuring unit was assessed based on a score of 1-3. For DE, a Score 1 was assigned when no DE or isolated small particles were present. Score 2 indicated that DE covered more than 30% of the canal walls and a Score 3 indicated that DE covered the entire canal wall. For the SL, a score 1 was assigned when a regular pattern of open dentinal tubules and no SL was present.

Score 2 indicated that some dentinal tubules were open and the others were covered by SL. Score 3 was assigned when a continuous SL covered the canal walls and no dentinal tubules were seen.

Table 2. Mean (SD) scores of smear layer removal at 1, 5 and 10 mm from the WL.

Statistical analysis

The average score of each level was calculated by dividing the sum of all individual scores by the number of evaluation units. Mean scores for DE and SL were then calculated for each tooth and for each group and statistically analysed for significance between groups using the Kruskal-Wallis non-parametric Anova and Dunn’s tests.

The results obtained at each evaluation level within each group were analysed using Friedman’s test and Turkey’s multiple comparison test. The level of significance was set at p<0.05.

Results

The inter-observer reproducibility was 94% which constituted a strong agreement. Therefore, the scoring of the samples was considered reliable. The mean scores of DE and SL recorded at 1, 5 and 10 mm from the WL are listed in subsequently Tables 1 and 2. Figure 1 is a representative image of scores 1, 2 and 3 for DE and SL respectively. In all groups, SEM observation showed cleaner canals at 5 and 10 mm compared to 1 mm from the WL.

Comparison between groups showed that at all evaluation levels the canals in which the irrigants were agitated with either Endoactivator, Eddy or XP-endo Finisher had significantly less surface DE and SL than those in which the control was used. XP-endo Finisher removed significantly more DE and SL than Endoactivator and Eddy, while no significant differences were observed between Endoactivator and Eddy. Based on these results, the null hypothesis was rejected.

Discussion

This study evaluated the cleanliness of root canals after the use of different irrigation/agitation systems. After root canal preparation, remnants of DE or the SL may prevent the adaptation of filling materials to the root canal walls, thus jeopardising their sealing properties. Therefore, DE and SL remnants should be totally removed before canal obturation.

As per protocol, a combination of NaOCl and EDTA solutions were used during root canal instrumentation, as well as for the final irrigation/agitation step. This combination of irrigants was found to be one of the more effective irrigation regimes. It produced clean root canal walls (Baumgartner and Mader, 1987; Hülsmann et al, 1997; Mayer et al, 2002).

However, several studies (Schafer and Zapke, 2000; Peters and Barbakow, 2000; Ahlquist et al, 2001) have demonstrated that none of the irrigants and irrigation techniques are totally efficient in completely cleaning the irregular areas of canal walls. For that reason, an additional irrigation/agitation step has been proposed (Duque et al, 2017). In this study Endoactivator, Eddy and XP-endo Finishing irrigation/agitation techniques were tested.

A conventional needle irrigation technique was used as control. This is because since the last, a no-agitation method, is still frequently used in endodontic treatment.

Previous studies

The results reported here are in support of Caron et al (2010), Jiang et al (2012), Kumar et al (2015), Mendonça et al (2015), Bayari (2017), demonstrating that improved DE and SL removal is accomplished with supplementary irrigation/agitation techniques.

Previous studies (Peters and Barbakow, 2000; Paque et al, 2006) have shown that regardless of the instrument or instrumentation technique, the most apical part of the root canal is the least clean. Findings by Abou-Rass and Piccinino (1982) have shown that instruments used for irrigation have to be in close proximity to the root canal walls in order for irrigants to be effective.

Present study

In the present study, the 30-gauge needle used for the final step of irrigation in canals prepared to a size 30/.07 did not allow the irrigants to reach the canal wall surfaces. In contrast, the action of Endoactivator, Eddy or XP-endo Finishing not only depends on irrigant agitation, but also on agitating action closer to the canal walls without removing dentine.

Endoactivator and Eddy generate intracanal agitation of irrigants though acoustic streaming and cavitation (Bayari et al 2017). At all evaluation levels, no significant differences were found between both sonic irrigation/agitation systems. They were significantly more effective than the control (a conventional needle irrigation technique).

On the other hand, other studies (Takeda et al, 1999; Saber and Hashem 2011) have shown that irrigation/agitation systems do not result in significant cleaning differences with the conventional needle irrigation technique. These findings are not supported by the current study or by others (Blank-Gonçalves et al, 2012).

Sonic agitation systems increase the effectiveness of the irrigating solutions. Especially in hard-to-reach areas such as the apical third. Also, in particular, in the area 1 mm from the WL. The difference between studies may be due to the use of different instruments and/or instrumentation techniques. As well as different experimental models and evaluation protocols.

Method used for observation

In the current study, the method used for observation of canal wall surfaces was similar to one used in a previous study (Zmener et al, 2005). This method allows for examination of both halves of the root. Thus examining the entire circumference of the root canal at a predetermined distance from the WL.

The XP-endo Finishing provided significantly cleaner canal walls at all evaluation levels. These results confirmed the claim of the manufacturer. This stated that the phase transformation of the alloy of the instrument when used at body temperature causes the file to expand. As a result, this improves the cleaning activity on wall irregularities.

It should be noted that previous in vitro studies did not use Endoactivator and Eddy at body temperature. (Khaord et al, 2010; Uroz-Torres, et al 2010; Bayari et al, 2017; Duque et al, 2017; Urban et al, 2017).

In this study, the final irrigation/agitation techniques were tested under similar temperature conditions. The objective was to standardise the experimental procedures. To mimic as much as possible the clinical conditions of the oral environment.

Conclusions

Within the limitations of this study, the results indicate that all three agitation devices, along with 5.25% NaOCl and 17% EDTA solution, as a final step in irrigation of root canals have greater potential for DE and SL removal when used in straight root canals.

The XP-endo Finisher was superior to Endoactivator, Eddy and a conventional needle irrigation technique. However, none of the tested techniques were able to completely clean the root canal walls.


For the list of references please contact [email protected].

This article first appeared in Endodontic Practice today. You can read the latest issue here. 